Possible troubles in the process of using desktop vacuum drying oven and its treatment

Fault phenomenon

Reasons

Approach

No power on boot

No power outlet

Check if the line is tripped and the outlet is good

The power cord plug is not plugged in or disconnected

Reinsert or repair the power cord

Fuse tube is not installed or broken

Check the vacuum box circuit no short circuit, more fuse tube (instrument power transformer short circuit, the heater short circuit, grounding short circuit causes the fuse to blow).

Instrument PV screen display

" □ □ □ □ ”

The temperature sensor Pt100 is bad

Check Pt100 for replacement ( 0 °C to 100 Ω , 0.3 Ω / °C )

Temperature sensor wiring disconnected

Reconnect

Instrument measurement range is wrong

Reset

Do not warm up

Low temperature setting

Set temperature SV ≥ RT+10 °C

RT is (ambient temperature)

The output circuit of the instrument is disconnected

Reconnect

Temperature controller has no output signal or bad or SCR bad

Exchange

Electric heater bad (open circuit, short circuit)

Exchange

Do not warm up

Enable timing or incorrect settings

ST=0 or ST= (heating + constant temperature) time

Temperature out of control or set temperature and measurement temperature error overshoot resulting in static error, overshoot

Temperature controller output is out of control

3041 or bad BTA , exchange

Does not meet the conditions of use

SV ≥ RT+10 °C

Pt sensor bad contact

Clear contact resistance

The relevant parameters are set incorrectly

Reset the relevant parameter example P, etc.

Measuring temperature and

The actual temperature in the box is too bad

No vacuum

Vacuum degree

Measure the mercury temperature sensor on the shelf

Reposition

Instrument or parameter changes

Re-correct SC , HL parameters

Can't vacuum

Vacuum pump selection model specifications are wrong

Should choose pumping rate not less than 2 liters per second

Various connection tubes (heads) are loose and the inner diameter is too small

Re-select the appropriate inner diameter connecting pipe, fastened at the joint

Vacuum table is bad

replace

The door is not closed

Adjust the buckle distance

Door seal rubber aging loses its elasticity

Replace door seal

Vent valve, vacuum valve position is wrong

Adjust position

Fault phenomenon

Reasons

Approach

Can't vacuum

Bad solenoid valve ( 6090 , 6210 )

Replace solenoid valve

Leakage

( 24 hours vacuum

From -0.1Mpa to

Greater than -0. 092Mpa )

Various connection pipe leaks

Replace after inspection

In addition to type 6050 , 6050B , 6051 and 6053 , the deformation of the heater “ O ” type seal ring is not compressed.

Tighten the heater base with the Allen key (on the back of the liner) or replace the " O " ring

The valve is not positioned correctly

Place the correct position

Vacuum valve leak

Exchange

Solenoid valve leaks tightly (for type 6090 , 6210 )

Exchange

Vacuum Forming Machine

Vacuum forming machine is a simplified version of thermoforming, where a sheet of plastic is heated to a forming temperature, stretched onto a single-surface mold, and forced against the mold by a vacuum. This process can be used to form plastic into permanent objects such as turnpike signs and protective covers. Normally draft angles are present in the design of the mold (a recommended minimum of 3°) to ease removal of the formed plastic part from the mold.